Dlab fixed volume micropipette
Dlab fixed volume micropipette’s volume range is 5μL-5000μL.
Description
This Dlab fixed volume micropipette has features as below:
- Light weight, ergonomic, low force design.
- Digital display clearly reads volume setting.
- Easy to calibrate and maintain with tool supplied.
- Design helps avoid repetitive strain injuries.
- Calibrated in accordance with lSO8655, Each pipette suppliedwith individual test certificate.
Specifications:
Volume Range | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | ||
---|---|---|---|---|
% | μL | % | μL | |
5μL | ±1.3 | ±0.065 | ±1.2 | ±0.06 |
10μL | ±0.8 | ±0.08 | ±0.8 | ±0.08 |
20μL | ±0.6 | ±0.12 | ±0.5 | ±0.1 |
25μL | ±0.5 | ±0.125 | ±0.3 | ±0.075 |
50μL | ±0.5 | ±0.25 | ±0.3 | ±0.15 |
100μL | ±0.5 | ±0.5 | ±0.3 | ±0.3 |
200μL | ±0.4 | ±0.8 | ±0.2 | ±0.4 |
250μL | ±0.4 | ±1.0 | ±0.2 | ±0.5 |
500μL | ±0.3 | ±1.5 | ±0.2 | ±1.0 |
1000μL | ±0.3 | ±3.0 | ±0.2 | ±2.0 |
2000μL | ±0.3 | ±6.0 | ±0.15 | ±3.0 |
5000μL | ±0.3 | ±15 | ±0.15 | ±7.5 |